Scroll Up Icon

Model Asphire AS-S3:-

Model Asphire AS-S3:-

Asphire Spectro Model AS-S3 is an Optimal Decision to the quick analysis with high accuracy and precision for the customer. These laboratory Instruments can work due to excitation of the electrons via the addition of enery which is in the form of some wavelength of the electromagnetic radiation which is in the form of some wavelength of the electromagnetic radiation which is usually light or ultraviolet.

AS-S3 is an advanced CCD/CMOS based Optical Emission Spectrometer having multi base structure with multiple matrices ( Ferrous, Aluminium, Copper, Nickel , Zinc .Tin & etc.) For the Analysis of chemical composition of the metal. AS-S3 is one of the smallest and lightest ultra-compact spark Emission, Excellent performance with competitive price. AS-S3 Model have 32+ or 55+ elements and we can also upgrade as per customer requirement.

Technical Specifications:-

Optical System-

  • Optical Air System
  • Stabilized against temperature fluctuation
  • Effective wavelength ranges: 115-800 nm, the applicable and configured wavelength range is based on the customer's application requirements.
  • Multi High Resolution CCD/CMOS system
  • High luminosity holographic grating
  • Argon purged optical chamber with Proprietary Low Consumption argon device.

Analysis Module:

  • Optimized to customer requirements.
  • Expansion of additional module on site possible.

Asphire AS-S3 Features:

  • Optical Systems
  • Wavelength Range: 115-800 nm |130-800 nm |140-800 nm
  • Focal Length: 300mm / 500mm / 750 mm
  • Holographic Grating: 1800 / 2400 / 3600 grooves/mm
  • CMOS / CCD detectors ranging from 2-16

Analysis Module:
  • Optimized to customer requirements.
  • Expansion of additional module on site possible.
Spark Stand :-
  • Open spark stand for high sample throughput and various kinds of sample geometries
  • Minimized argon consumption controlled with software
  • Optimized argon flow
  • Reduced cleaning intervals
  • Spark stand plate easily exchangeable.
Excitation System:-

Fully digitalized plasma generator with digital discharge definition, digital pulse generator and digital offline pulse control., Max spark frequency 1000Hz.

Readout System:

Single board readout system dedicated controlled system for each CMOS/CCD detectors.

Environmental Condition:

  • Room Temperature should be 22 ºC (± 5 ºC). | 18-28 ºC
  • Temperature maintained with Air condition (AC)
  • Relative Humidity < 80%.
  • Atmosphere free of corrosive vapors and high dust pollution

Electrical Power Supply :

  • 230 V ± 10 V (AC) | 50 Hz. |
  • Earthing Voltage < 2 V.
  • Power supply is singe phase.
  • Power rating Max During spark 800 VA.